VAC-V2 PERME Gas Permeability Tester

Measures how gas permeable a membrane is. Capable of measuring the transmission rate of permanent gases, air, oxygen, nitrogen and carbon dioxide - particularly useful for the permeability of carbon dioxide. Three test chambers.

The PERMETM VAC-V2 measures the gas permeability of   a membrane.

Capable of measuring the transmission rate permanent gases, air, oxygen, nitrogen and carbon dioxide and is particularly useful for the permeability of CO2.

The VAC-2 has three test chambers to ensure a quick test time, is very sensitive and highly linear, with a measuring range of 0.05 – 50,000 cm3/m2/day. Software calculates rate of transmission, solubility and diffusion coefficients.

Standards
Complies with: ISO 2556, ISO 15105-1, ASTM D1434, GB 1038, JIS K7126-A

Related products
The PERME VAC-V2 replaces the Lyssy L100-5000 manometric gas permeability tester, which has been discontinued by PBI-Dansensor.

Also in the PERME series:

Features

  • Uses a Manometric principle
  • Three independent test chambers, can display individual cell data and statistics for the three as a group
  • High precision vacuum sensor to ensure consistent testing
  • User selectable temperature control device for testing between 5 – 95 °C
  • Optional dual test modes – conventional and non-standard
  • Powerful software for automatic calculation of solubility, permeability and diffusion coefficients
  • Microsoft ® Excel ® format conversion and graphical display
  • Software can calculate gas transmission rate, diffusion coefficient, solubility coefficient and permeability coefficient at any temperature (-273 – +200 °C )
  • Reliable performance, nonstop operation and improved test efficiency
  • Easy calibration of reference film

Principle
A preconditioned test specimen is clamped between the upper chamber and the lower chamber of a test cell. A vacuum is applied to the low-pressure lower chamber and then to the whole system.

When the specified degree of vacuum is reached, the lower test chamber is shut off and test gas of a selected pressure is fed into the upper high pressure test chamber.

Ensuring that a constant differential pressure (adjustable) is maintained across the specimen, under the gradient of differential pressure the test gas permeates from the high-pressure side to the low-pressure side.

The pressure in the low-pressure side is monitored in real time and the rate of transmission is calculated.

Technical Specifications

Test range
0.05 to 50.000 cm3 / m2 / day   0.1 MPa (normal)
Extendable to   500.000 cm3/m2 / day 0.1 MPa (extension volume)
Normal and extension volume can be selected using extension vacuum accessory.
Temp. control range
5 ºC – 95 ºC (water medium)
Temp. control accuracy
+/- 0.1 ºC
Humidity control range
0 %RH, 2 %RH – 98.5 %, 100 %RH (additional humidity generator is required)
Humidity control accuracy
+/- 1 %RH
Vacuum resolution
0.1 Pa
Vacuum degree
< 20 Pa
Test gas
O2, N2, CO2
Test pressure
-0.1 MPa – +0.1 MPa
Pressure of gas source
0.4 MPa – 0.6 MPa
Specimen size
diameter –   97 mm
Transmission area
38.48 cm ² (diameter 70 mm)
No. of specimens
3 pieces

Physical specifications

  • Dimensions: 760 x 575 x 450 mm (L x B x H)
  • Weight: 88 kg
  • Power: AC 220V 50Hz / 60Hz

Includes

  • Base unit
  • Software
  • Extension vacuum accessory
  • Sample cutter
  • Vacuum pump
  • Vacuum grease
  • Filter paper

Optional

  • Blades
  • Oil for vacuum pump
  • Humidity generator

Note
Compressed air are not provided