Seal Integrity Testing Devices
For testing the seal integrity of flexible packaging.
These products are used to test the seal integrity of a flexible package. The creep test function checks the package seal under sustained pressure. The creep-to-burst test function sequentially checks minimum required seal strength and maximum pressure at burst. Tests can also simulate high altitude conditions.
Applications
Flexible package testing for Sterilized Medical Devices, Food and Pharmaceuticals.
Specifications
Model 2600 Seal Integrity Tester
Checks the seal integrity on a wide range of porous and non-porous flexible packages using a pressurized air burst test. Open and closed package test fixtures are available in a variety of standard and custom configurations. Pressurized burst test uses ASTM 1140 method.
Model 2700 Integrity Tester
Detects leaks in non-porous packages with a vacuum test chamber. Leak location can be identified by submerging the sample in water.
Features
- Model 2600 Seal Strength Tester
- RS 232 data acquisition interface
- Automatic Pressure Control
- 0.2% full scale pressure calibration accuracy
Connections
Please specify voltage requirements when ordering
Package Test System Specifications
Vacuum Source
- Use vacuum pump 13-26-01 or equivalent
- Electrical Power 85-264 VAC 47-63 Hz
Mechanical
- Weight: Control Console 11.4 lbs (5.2 kg)
- Weight: Vacuum Chamber: Approximately 3 lbs (1.4 kg)
- Parallel Printer Port: Standard db 25 female connector, Centronics format
Environmental
Temperature Range
- Operating: 40 °F – 110 °F (4 °C – 43 °C)
- Non-operating: 0 °F – 125 °F (-184 °C – 52 °C)
Humidity: Up to 90% R.H.
Performance
| Vacuum Accuracy | 0.2% full scale 29.9 in. Hg | ||
| Testing Range | mm Hg | ln Hg | Altitude |
| Absolute Pressure (max) | 760 | 29.9 | N/A |
| Absolute Pressure (min) | 1 | 1 | N/A |
| Gauge Pressure (max) | 760 | 29.9 | N/A |
| Gauge Pressure (min) | 1 | 1 | N/A |
| Altitude (max) | N/A | N/A | 50,000 |
| Altitude (min) | N/A | N/A | depends on geographic location |
Automatic Control Console
Vacuum Source
70-100 PSIG – Dry, oil-free, instrument quality air
Electrical Power: 85-264 VAC 47-63 Hz
Mechanical
Weight: 13 lbs (5.9 kg)
Serial (RS232) Output Port: Standard DB 9 female connector
Parallel Printer Port: Standard db 25 female connector, Centronics format
Environmental
Temperature Range
Operating :40 °F – 110 °F (4 °C – 43 °C)
Non-operating: 0 °F – 125 °F (-184 °C – 52 °C)
Humidity: Up to 90% R.H.
Preformance
Pressure Accuracy: 0.2% full scale (50 PSI)
Pressure Resolution: 0.3″ H20
Sensitivity Adjustments: 9 Programmable Settings
Flow Rate Adjustments: 9 Programmable Settings
Ordering Information
Model 2600 Catalog Number 13-25
Model 2700 Catalog Number 13-26
Sensing Probes:
| Part No. | Description | Weight |
| 13-25-05 | Closed Package Test Fixture with 3/16 in. Diameter Needle Probe Assembly | 2.7 kg (6.0 lbs) |
| 13-25-07 | Closed Package Test Fixture with Double Needle Probe Assembly |
Open Package Test Fixture:
| Part No. | Description | Weight |
| 13-25-01 | 12″ Open Package Test Fixture | 3.2 kg (7 lbs) |
| 13-25-02 | 24″ Open Package Test Fixture | 10.4 kg (23 lbs) |
Automatic Control Console:
| Part No. | Power Supply | Printer Output | RS232 Output Port | Data Export |
| 13-25-00-0001 | 110 VAC; 60 Hz | Yes | Yes | Yes |
| 13-25-00-0002 | 220 VAC; 50 Hz | Yes | Yes | Yes |